All issues > Volume 25(8); 1982
- Original Article
- J Korean Pediatr Soc. 1982;25(8):787-790. Published online August 31, 1982.
- The Evaluation of AO Bilirubinometer for Neonatal Hyperbilirubinemia.
- Hae Seong Kim, Joongh Gon Kim, Keun Lee
- 1Dept. of Pediatrics, College of medicine, Ewha Womans University, Korea.
- Abstract
- AO bilirubinometer is a direct reading spectrophotometer specifically designed to measure total bilirubin using sera from newborn infantsl A study was performed to compare the bilirubin values obtained using the AO bilirubibometer and modified Diazo method(Rappaport method). The results were as follows : 1. The average percentage of recovery of AO bilirubinometer was 98.7%, while modified Diazo method was represented as 95.3%. 2. In comparison of two bilirubin techniques, the regression is linear, expressed by Y=1.16X-2.42, where values obtained as the bilirubinometer was on X axis and those from the modified Diazo mtethod on Y axis. The correlation coefficient was 0.95. 3. The variability of bilirubin values by cuvette was negligible, average percentage of variation coefficient being 0.1%. 4. The variability of bilirubin values by examiner was also negligible, the evarage percentage of bilirubin variation coefficient being 0.67%.
Keywords :Direct reading sphectrophotometer;Modified Diazo method;Correlation coefficient