J Korean Pediatr Soc. 1982;25(8):787-790. Published online August 31, 1982.
The Evaluation of AO Bilirubinometer for Neonatal Hyperbilirubinemia.
Hae Seong Kim, Joongh Gon Kim, Keun Lee
1Dept. of Pediatrics, College of medicine, Ewha Womans University, Korea.
Abstract
AO bilirubinometer is a direct reading spectrophotometer specifically designed to measure total bilirubin using sera from newborn infantsl A study was performed to compare the bilirubin values obtained using the AO bilirubibometer and modified Diazo method(Rappaport method). The results were as follows : 1. The average percentage of recovery of AO bilirubinometer was 98.7%, while modified Diazo method was represented as 95.3%. 2. In comparison of two bilirubin techniques, the regression is linear, expressed by Y=1.16X-2.42, where values obtained as the bilirubinometer was on X axis and those from the modified Diazo mtethod on Y axis. The correlation coefficient was 0.95. 3. The variability of bilirubin values by cuvette was negligible, average percentage of variation coefficient being 0.1%. 4. The variability of bilirubin values by examiner was also negligible, the evarage percentage of bilirubin variation coefficient being 0.67%.

Keywords :Direct reading sphectrophotometer;Modified Diazo method;Correlation coefficient