Oscillometric Blood Pressure of Upper and Lower Extremities and Aortic Blood Flow Velocity in Neonates

Journal of the Korean Pediatric Society 2000;43(4):484-488.
Published online April 15, 2000.
Oscillometric Blood Pressure of Upper and Lower Extremities and Aortic Blood Flow Velocity in Neonates
Sang Young Bae, Hae Soon Kim, Se Jung Sohn, Young Mi Hong
Department of Pediatrics, College of Medicine, Ewha Womans University, Seoul, Korea
Oscillometry에 의한 정상 신생아의 상하지 혈압 측정과 대동맥 혈류 속도에 관한 연구
배상영, 김혜순, 손세정, 홍영미
이화여자대학교 의과대학 소아과학교실
Abstract
Purpose
: We studied the blood pressure difference between the upper and lower extremities in healthy newborn infants and the effect of isthmic narrowing of the aorta on making a possible difference.
Methods
: The blood pressure was measured with an oscillometric blood pressure device from four extremities of 62 healthy infants aged 1-15 days. A Doppler echocardiography was performed for each infant to measure the aortic blood flow velocities in the ascending, just above and below the aortic arch isthmic narrowing.
Results
: The blood pressure readings were as follow : right arm 69.7(¡¾7.5)/42.7(¡¾7.2)mmHg, right calf 69.5(¡¾6.1)/42.6(¡¾6.3)mmHg, left arm 69.0(¡¾7.3)/43.1(¡¾6.3)mmHg, and left calf 68.9(¡¾7.4)/42.9(¡¾5.6)mmHg. The estimated pressure gradient between the ascending aorta and aorta below the isthmus was 2.4¡¾1.3mmHg and between opposite sides of the isthmus was 1.2¡¾0.8 mmHg.
Conclusion
: Unlike in childhood and adolescence, the blood pressure in the lower extremities of healthy newborn infants is not higher than in the upper extremities. But the isthmic narrowing of the aortic arch does not explain this phenomenon.
Key Words: Blood pressure, Oscillometric method, Newborn infant, Doppler echocardiography